Measurement Specifications
|
Relative Humidity(RH)
|
Range:0-100%RH(0.00-1.00)
Resolution:0.10% RH
Accuracy:Sensormeasurement accuracy is variable across a range of RH
Equilibration Time (r,63%):< 25 s (response time in 1 m/s air stream)
Hysteresis: ±0.80% RH,typical
Long-term Drift: ±0.25% RH/year, typical
|
Temperature
|
Range:-40.00 - 80.00°C
Resolution:0.1°C
Accuracy:+0.20℃
Equilibration Time(t.63%):< 165s(response time in 1 m/s air stream
Long-term Drift:< 0.03 °C/ year, typical
|
Vapor Pressure
|
Range:0-47kPa
Resolution:0.01 kPa
Accuracy:Sensor measurement accuracy is variable across a range of temperatures and RH.
|
Barometric Pressure
|
Range:1-120 kPa
Resolution:0.01 kPa
Accuracy:+0.05 kPa at 25°C
Equilibration Time(t,63%):< 10ms
Long-term Drift:< 0.1 kPa/year, typical
|
Communication Specifications
|
Output
|
DDI serial or SD1-12 communication
|
Data Logger Compatibility
|
Anydata acquisition system capable of 4.0- to 15.0-VDC power and serial or SDI-12 communication.
|
Physical Specifications
|
Dimensions
|
Diameter: 2.0 cm (0.8 in)
Height: 5.4 cm (2.1 in)
|
Operating Temperature Range
|
Minimum:-40.00°C
Maximum:80.00°C
NOTE:Sensors may be used at higher temperatures under certain conditions:contact Customer Support for assistance.
|
Cable Length
|
5 m(standard)
75 m (maximum custom cable length)
NOTE:Contact Customer Support if a nonstandard cable lengthis needed.
|
Cable Diameter
|
0.165 +0.004 in (4.20 +0.10 mm), with minimum jacket of0.030in(0.76 mm)
|
Connector Types
|
3.5-mm stereo plug connector or stripped and tinned wires
|
Conductor Gauge
|
22-AWG/24-AWG drain wire
|
Electrical and Timing Characteristics
|
Supply Voltage (VCC to GND)
|
Minimum:4.0 VDC
Maximum:15.0 VDC
|
Digital Input Voltage(Logic High)
|
Minimum:2.8V
Typical: 3.6 V
Maximum:5.0 V
|
Digital Input Voltage(Logic Low)
|
Minimum:-0.3V
Typical: 0.0 V
Maximum:0.8 V
|
Digital Output Voltage(Logic High)
|
Minimum: NA
Typical:3.6 V
Maximum: NA
|
Power Line Slew Rate
|
1.0 V/ms
|
Current Drain (During Measurement)
|
Minimum:1.2 mA
Typical: 3.5 mA
Maximum:12.0 mA
|
Current Drain (While Asleep)
|
Minimum:NA
Typical:0.03 mA
Maximum: NA
|
Power Up Time(DDI Serial)
|
Minimum:NA
Typical: NA
Maximum:110 ms
|
Power Up Time(SDI-12)
|
Minimum:140 ms
Typical:260 ms
Maximum:400 ms
|
Power Up Time(SDI-12, DDI Serial Disabled)
|
Minimum: NA
Typical:170 ms
Maximum:200 ms
|
Measurement Duration
|
Minimum: NA
Typical:50 ms
Maximum:NA
|
Other
|
Compliance
|
Manufactured under ISO 9001:2015
2004/108/EC and 2011/65/EU
EN61326-1:2013
EN50581:2012
|